MEMORANDUM OPINION
McKEAGUE, District Judge.
This case presents four claims growing out of injury caused by a malfunctioning pacemaker. The pacemaker had been surgically implanted in plaintiff Brett Strong on July 16, 1992 and was removed on February 16, 1993. During this period, Mr. Strong experienced inconsistent and inappropriate heart rates, allegedly resulting in lost wages, additional medical expenses, pain and suffering, and anxiety. In count I of the complaint, he alleges defendant Teleetronics Pacing Systems, Inc. (“Teleetronics”), is liable for having negligently manufactured the pacemaker. In counts II and III, he alleges defendant Thomas S. Martin, an employee or agent of Teleetronics, negligently evaluated the pacemaker’s functioning and negligently misrepresented that it was functioning properly. Count IV contains the claim of Mr. Strong’s wife, Lauri Strong, for loss of consortium.
The complaint was originally filed in the Kent County Circuit Court. Defendants removed the action to this Court primarily for the reason that plaintiffs’ negligent manufacture claim asserts a wrong remediable exclusively under federal law. Now before the Court are plaintiffs motion to remand and defendants’ motion to dismiss and for summary judgment.
I. PLAINTIFFS’ MOTION TO REMAND
Normally, a cause of action arises under federal law only when a plaintiffs well-pleaded complaint raises issues under federal law.
Metropolitan Life Ins. Co. v. Taylor,
481 U.S. 58, 63, 107 S.Ct. 1542, 1546, 95 L.Ed.2d 55 (1987). Count I of the complaint does not expressly state a claim arising under federal law. Defendants correctly assert, however, that where Congress has so thoroughly occupied a legislative field as to leave no additional room for the states to regulate, any claim arising in the field is necessarily federal in character.
Id.,
at 63-64, 107 S.Ct. at 1546^7.
The subject Teleetronics pacemaker is a Class III Medical Device regulated under the Medical Device Amendments (“MDA”) to the Federal Food, Drug and Cosmetic Act (“FDCA”), 21 U.S.C. § 321
et seq.
The MDA contains an express preemption provision, 21 U.S.C. § SdOkia),
which has been consistently construed as preempting state law product liability claims involving the safety or effectiveness of Class III devices.
Mendes v. Medtronic, Inc.,
18 F.3d 13 (1st Cir.1994);
Stamps v. Collagen Corp.,
984 F.2d 1416 (5th Cir.1993),
cert. denied,
— U.S. -, 114 S.Ct. 86, 126 L.Ed.2d 54 (1993);
King v. Collagen Corp.,
983 F.2d 1130 (1st Cir.1993),
cert. denied,
— U.S. -, 114 S.Ct. 84, 126 L.Ed.2d 52 (1993);
Brown v. Medtronic, Inc.,
852 F.Supp. 717 (S.D.Ind.1994);
Bravman v. Baxter Healthcare Corp.,
842 F.Supp. 747 (S.D.N.Y.1994);
Griffin v. Medtronic, Inc.,
840 F.Supp. 396 (D.Md.1994);
Kemp v. Pfizer, Inc.,
835 F.Supp. 1015 (E.D.Mich.1993). In fact, two of the cited eases specifically held that claims for defectively manufactured pacemakers are preempted.
Mendes, supra; Griffin, supra.
Plaintiffs have presented no persuasive reason to depart from the rule established by these authorities. They argue the defective manufacture claim may be characterized as alleging the pacemaker was an “adulterated device,” a claim exempt from preemption under 21 C.F.R. § 808.1(d)(6)(ii).
Even assuming a defectively manufactured pacemaker may be deemed an “adulterated device,” as defined at 21 U.S.C. § 351(h),
plaintiffs have failed to show that a violation of the MDA “good manufacturing practice requirements” prescribed at 21 U.S.C. § 360j(f)(l),
rendering the pacemaker adulterated, is actionable under Michigan law. They have not done so because they cannot do so. “Violations of the FDCA do not create private rights of action.”
Gile v. Optical Radiation Corp.,
22 F.3d 540, 544 (3rd Cir. 1994). “Only the government has a right to take action with respect to adulterated products.”
Id.
Thus, to characterize a defective manufacture claim as an “adulterated device” claim does not avoid preemption.
Id.
The state law claim contained in count I of the complaint relates to the safety or effectiveness of a medical device subject to regulation under the MDA. The claim is preempted by the MDA. As such, the claim presents a question of federal law, establishing federal court jurisdiction. 28 U.S.C. § 1331. Removal was therefore proper under 28 U.S.C. § 1441(a) and plaintiffs’ motion for remand must be denied.
II.DEFENDANT TELECTRONICS’ MOTION TO DISMISS
The above analysis is also dispositive of defendant Telectronics’ motion to dismiss under Fed.R.Civ.P. 12(b)(6). Because plaintiffs’ count I state law tort claim arises in a field of regulation completely occupied by the MDA, and because allowing plaintiffs to prosecute their claim would have the effect of permitting enforcement of a requirement relating to the safety or effectiveness of the pacemaker — a requirement different from and in addition to those applicable under the MDA — it is preempted pursuant to 21 U.S.C. § 360k(a) and must be dismissed.
III.DEFENDANT MARTIN’S MOTION FOR SUMMARY JUDGMENT
Defendant Martin moves for summary judgment under Fed.R.Civ.P. 56 contending there is no genuine issue as to any material fact and that he is entitled to judgment as a matter of law. The motion is made on the strength of Mr. Martin’s own affidavit, indicating that when he monitored the Telectronics pacemaker implanted in Brett Strong, it was operating correctly and that he made only accurate representations about its functioning.
The standards governing the motion are well established. They are well summarized by the Sixth Circuit in
Mitchell v. Toledo Hospital,
Free access — add to your briefcase to read the full text and ask questions with AI
MEMORANDUM OPINION
McKEAGUE, District Judge.
This case presents four claims growing out of injury caused by a malfunctioning pacemaker. The pacemaker had been surgically implanted in plaintiff Brett Strong on July 16, 1992 and was removed on February 16, 1993. During this period, Mr. Strong experienced inconsistent and inappropriate heart rates, allegedly resulting in lost wages, additional medical expenses, pain and suffering, and anxiety. In count I of the complaint, he alleges defendant Teleetronics Pacing Systems, Inc. (“Teleetronics”), is liable for having negligently manufactured the pacemaker. In counts II and III, he alleges defendant Thomas S. Martin, an employee or agent of Teleetronics, negligently evaluated the pacemaker’s functioning and negligently misrepresented that it was functioning properly. Count IV contains the claim of Mr. Strong’s wife, Lauri Strong, for loss of consortium.
The complaint was originally filed in the Kent County Circuit Court. Defendants removed the action to this Court primarily for the reason that plaintiffs’ negligent manufacture claim asserts a wrong remediable exclusively under federal law. Now before the Court are plaintiffs motion to remand and defendants’ motion to dismiss and for summary judgment.
I. PLAINTIFFS’ MOTION TO REMAND
Normally, a cause of action arises under federal law only when a plaintiffs well-pleaded complaint raises issues under federal law.
Metropolitan Life Ins. Co. v. Taylor,
481 U.S. 58, 63, 107 S.Ct. 1542, 1546, 95 L.Ed.2d 55 (1987). Count I of the complaint does not expressly state a claim arising under federal law. Defendants correctly assert, however, that where Congress has so thoroughly occupied a legislative field as to leave no additional room for the states to regulate, any claim arising in the field is necessarily federal in character.
Id.,
at 63-64, 107 S.Ct. at 1546^7.
The subject Teleetronics pacemaker is a Class III Medical Device regulated under the Medical Device Amendments (“MDA”) to the Federal Food, Drug and Cosmetic Act (“FDCA”), 21 U.S.C. § 321
et seq.
The MDA contains an express preemption provision, 21 U.S.C. § SdOkia),
which has been consistently construed as preempting state law product liability claims involving the safety or effectiveness of Class III devices.
Mendes v. Medtronic, Inc.,
18 F.3d 13 (1st Cir.1994);
Stamps v. Collagen Corp.,
984 F.2d 1416 (5th Cir.1993),
cert. denied,
— U.S. -, 114 S.Ct. 86, 126 L.Ed.2d 54 (1993);
King v. Collagen Corp.,
983 F.2d 1130 (1st Cir.1993),
cert. denied,
— U.S. -, 114 S.Ct. 84, 126 L.Ed.2d 52 (1993);
Brown v. Medtronic, Inc.,
852 F.Supp. 717 (S.D.Ind.1994);
Bravman v. Baxter Healthcare Corp.,
842 F.Supp. 747 (S.D.N.Y.1994);
Griffin v. Medtronic, Inc.,
840 F.Supp. 396 (D.Md.1994);
Kemp v. Pfizer, Inc.,
835 F.Supp. 1015 (E.D.Mich.1993). In fact, two of the cited eases specifically held that claims for defectively manufactured pacemakers are preempted.
Mendes, supra; Griffin, supra.
Plaintiffs have presented no persuasive reason to depart from the rule established by these authorities. They argue the defective manufacture claim may be characterized as alleging the pacemaker was an “adulterated device,” a claim exempt from preemption under 21 C.F.R. § 808.1(d)(6)(ii).
Even assuming a defectively manufactured pacemaker may be deemed an “adulterated device,” as defined at 21 U.S.C. § 351(h),
plaintiffs have failed to show that a violation of the MDA “good manufacturing practice requirements” prescribed at 21 U.S.C. § 360j(f)(l),
rendering the pacemaker adulterated, is actionable under Michigan law. They have not done so because they cannot do so. “Violations of the FDCA do not create private rights of action.”
Gile v. Optical Radiation Corp.,
22 F.3d 540, 544 (3rd Cir. 1994). “Only the government has a right to take action with respect to adulterated products.”
Id.
Thus, to characterize a defective manufacture claim as an “adulterated device” claim does not avoid preemption.
Id.
The state law claim contained in count I of the complaint relates to the safety or effectiveness of a medical device subject to regulation under the MDA. The claim is preempted by the MDA. As such, the claim presents a question of federal law, establishing federal court jurisdiction. 28 U.S.C. § 1331. Removal was therefore proper under 28 U.S.C. § 1441(a) and plaintiffs’ motion for remand must be denied.
II.DEFENDANT TELECTRONICS’ MOTION TO DISMISS
The above analysis is also dispositive of defendant Telectronics’ motion to dismiss under Fed.R.Civ.P. 12(b)(6). Because plaintiffs’ count I state law tort claim arises in a field of regulation completely occupied by the MDA, and because allowing plaintiffs to prosecute their claim would have the effect of permitting enforcement of a requirement relating to the safety or effectiveness of the pacemaker — a requirement different from and in addition to those applicable under the MDA — it is preempted pursuant to 21 U.S.C. § 360k(a) and must be dismissed.
III.DEFENDANT MARTIN’S MOTION FOR SUMMARY JUDGMENT
Defendant Martin moves for summary judgment under Fed.R.Civ.P. 56 contending there is no genuine issue as to any material fact and that he is entitled to judgment as a matter of law. The motion is made on the strength of Mr. Martin’s own affidavit, indicating that when he monitored the Telectronics pacemaker implanted in Brett Strong, it was operating correctly and that he made only accurate representations about its functioning.
The standards governing the motion are well established. They are well summarized by the Sixth Circuit in
Mitchell v. Toledo Hospital,
964 F.2d 577, 581-82 (6th Cir.1992), and need not be reiterated here.
Plaintiffs have presented no evidence rebutting defendant Martin’s affidavit. They contend they have had no opportunity to develop such evidence as discovery has been held in abeyance pending resolution of the instant motions.
Indeed, defendant’s motion, brought before plaintiffs have had a reasonable opportunity to conduct discovery, is premature. See
id.,
at 582;
Street v. J.C. Bradford & Co.,
886 F.2d 1472, 1478 (6th Cir.1989). The motion must be denied, without prejudice.
IV.REMAND TO STATE COURT
The above rulings establish (1) that the legitimate basis for federal subject matter jurisdiction, upon which the action was properly removed to this Court, has been extinguished; and (2) that further proceedings are required to resolve the state law
claims presented in counts II, III and IV. Under these circumstances, the Court, in the exercise of its discretion, declines to exercise continuing supplemental jurisdiction over the surviving state law claims. See 28 U.S.C. § 1367(c)(3);
Landefeld v. Marion General Hospital, Inc.,
994 F.2d 1178, 1182 (6th Cir. 1993). Accordingly, the pendent state law claims will be remanded to the Kent County Circuit Court for further proceedings.
An order consistent with this opinion shall issue forthwith.