In this prosecution for driving while intoxicated (Vehicle and Traffic Law § 1192 [2], [3]), the court has conducted a hearing to determine whether the Intoxilyzer 5000 is sufficiently reliable to permit the admissibility of its test results. The Intoxilyzer 5000 is used to determine the blood alcohol concentration of the defendant’s blood.

The expert witnesses and their backgrounds are summarized as follows:

[764]*764James Witler is the coinventor of the Intoxilyzer 5000 and the director of engineering for CMI, Inc. of Minturn, Colorado (the wholly owned subsidiary of Federal Signal Corporation of Oakbrook, 111., the manufacturer of the Intoxilyzer 5000). Mr. Witler planned, directed and coordinated the engineering, research and development of the machine, and supervises the repair of existing products.

Dr. Arthur Flores is the project director for the Office of Alcohol Countermeasures for the National Highway Traffic Safety Administration in Cambridge, Massachusetts. He holds a doctorate in physical chemistry, and his principal responsibility with respect to the Intoxilyzer 5000 is to conduct tests, both those that were established pursuant to the National Bureau of Standards, and other tests of a more informal nature, in an effort to determine the reliability of the Intoxilyzer 5000.

Dr. Leo Dalcortivo of the office of the Medical Examiner of Suffolk County holds the positions of director of laboratory services and chief toxicologist. Associated with that office since 1960, Dr. Dalcortivo was formerly the senior toxicological chemist and holds a doctorate in chemistry.

The final witness was Police Officer James McCarthy, who has been with the Suffolk County Police Department for 15 years and is currently with the breath test section of the department. He is the technical supervisor of the unit and teaches the operators how to use the breath test equipment. He also checks the instrument to ascertain if the instrument is working properly, turns the instrument on, and does the diagnostic checks. In general, his unit handles the calibration of the breath testing equipment.

Based on the testimony of these witnesses, the court makes the following findings regarding the workings and reliability of the Intoxilyzer 5000:

WORKINGS OF THE INTOXILYZER 5000

The Intoxilyzer 5000 uses an infra-red light beam passed through a chamber containing a sample of the subject’s breath in order to determine the concentration of ethyl alcohol in that subject’s blood. When the machine is turned on, it automatically purges the sample chamber, heats to 45 degrees centigrade, and indicates its readiness. The operator pushes a button and inserts a card. By way of control, fresh air is introduced and analyzed to obtain a reference for the test of the subject’s breath. Next, the subject blows into the sample [765]*765chamber, which holds 81 cubic centimeters of air and is bounded by two lenses. He must blow for four seconds at a sufficient pressure to insure that there is no "mouth alcohol” that could be detected by the machine. The slope of alcohol concentration in the breath is highest at the beginning stages of the test. Thereafter, the slope levels off and approaches 0% of change.

The breath analysis, which occurs once the slope has leveled off, involves the principle of "infra-red spectroscopy.” An infra-red source lamp projects infra-red rays directly through the sample chamber, where they interact with the subject’s breath sample. The rays are then channeled through a filter wheel which filters out all hydrocarbons which are not the subject of the test, to wit, everything except ethyl alcohol. An infra-red energy detector then picks up the infra-red rays. A reading of the alcohol concentration appears on a digital display terminal and is printed on the card, and the machine automatically purges itself. As a further control, air from a connected simulator is then introduced into the chamber and tested. Once the result is printed on the card, the machine again purges itself. This entire procedure takes about 95 seconds.

CHEMICAL TEST OR CHEMICAL ANALYSIS

Section 1192 (2) of the Vehicle and Traffic Law speaks of the chemical analysis of a person’s blood, whereas Vehicle and Traffic Law § 1194 speaks of the chemical test of a person’s breath, blood, urine or saliva for the purpose of determining alcohol or drug content of his blood. It would appear that these terms have been used synonymously, and, therefore, this court will treat them as if they are synonymous.

There is a threshold question raised with regard to whether the Intoxilyzer 5000 testing procedure constitutes a chemical test/chemical analysis as contemplated by the New York State Legislature. In support of a bill previously submitted in the New York State Senate, the observation is made that, "The Breathalyzer involves test solutions and a 'chemical test’ of the driver’s breath. Currently, police departments are purchasing and are using various instruments which employ techniques such as infra-red analysis rather than chemical techniques. These new instruments, such as the CMI [I]ntoxilyzer and the Smith and Wesson Model 2000, are extremely accurate and reliable. Courts have construed these new instruments as not involving 'chemical tests’ under sections 1194 [766]*766and 1195 of the [Vjehicle and [T]raffic [L]aw.” (Sen Kehoe, mem in support, S 3106, 1987-1988 Regular Sess.)

Moreover, courts of other States have held that the results of various Intoxilyzer models are admissible as a chemical test/chemical analysis (see, State v Moore, 307 A2d 548 [Del Super Ct]; Dollar v State, 287 Ark 153, 697 SW2d 93; Curry v State, 522 So 2d 887 [Fla Dist Ct App]), and the undisputed testimony at the hearing before this court was that a chemical reaction indeed occurs in the sample chamber of the Intoxilyzer 5000 when the infra-red rays interact with the breath sample. The molecules of alcohol absorb the energy of the infra-red radiation. Consequently, this court finds that the Intoxilyzer 5000 testing procedure is a chemical test/chemical analysis within the meaning of Vehicle and Traffic Law § 1194 (see also, People v Jones, 118 Misc 2d 687 [Intoximeter 3000]).

RELIABILITY OF THE INTOXILYZER 5000

The Intoxilyzer 5000 has appeared on the Conforming Products List of Evidential Breath Measurement Devices of the United States Department of Transportation/National Highway Traffic Safety Administration. This list has been adopted by the Commissioner of the New York State Department of Health.1

In 1984, the United States Department of Transportation promulgated rules and regulations for the testing of machines that analyze the alcoholic content of the breath. A manufacturer is required to submit at least one unit of its equipment to the offices of the National Highway Traffic Safety Administration in Cambridge, Massachusetts. The manufacturer is responsible for insuring that the machine is properly calibrated and operating properly. The machine is tested by a BASS method, that is, a breath alcohol sample simulator, which insures that there is a uniformity for all samples tested.

The testing is conducted at .05%, .101% and .151% blood alcohol content (hereinafter BAG), the Department of Transportation noting that no State has a higher BAG level than .15% for the determination of intoxication.

Free access — add to your briefcase to read the full text and ask questions with AI

People v. Summa, 140 Misc. 2d 763, 531 N.Y.S.2d 993, 1988 N.Y. Misc. LEXIS 471 (N.Y. Super. Ct. 1988).

140 Misc. 2d 763 (People v. Summa) — published by Counsel Stack Legal Research, free access to 12M+ legal documents.

Related

People v. Burnet
25 Misc. 3d 307 (New York Supreme Court, 2009)
People v. Cialino
14 Misc. 3d 999 (Criminal Court of the City of New York, 2007)
People v. Davis
8 Misc. 3d 158 (New York Supreme Court, 2005)
Government of the Virgin Islands v. Carela
44 V.I. 11 (Supreme Court of The Virgin Islands, 2001)
People v. Holmes
171 Misc. 2d 962 (Brighton Justice Court, 1997)
People v. Hampe
181 A.D.2d 238 (Appellate Division of the Supreme Court of New York, 1992)
People v. Drake
152 Misc. 2d 441 (Nassau County District Court, 1991)