21 CFR · Food and Drugs

§ 892.2055 — Radiological machine learning-based quantitative imaging software with predetermined change control plan.

eCFR · current through Jul 23, 2026

§ 892.2055 Radiological machine learning-based quantitative imaging software with predetermined change control plan.

(a)Identification. A radiological machine learning based quantitative imaging software with predetermined change control plan is a software-only device which employs machine learning algorithms on radiological images to provide quantitative imaging outputs. The device includes functions to support outputs such as view selection, segmentation and landmarking. The design specifications include planned modifications that may be made to the device consistent with an established predetermined change control plan.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Design verification and validation must include:
(i)A detailed descri