21 CFR · Food and Drugs
§ 886.4320 — Corneal storage medium with preservatives including antifungals.
eCFR · current through Jul 20, 2026
§ 886.4320 Corneal storage medium with preservatives including antifungals.
(a)Identification. Corneal storage medium with preservatives including antifungals is a device that is used to temporarily preserve human cornea tissue between harvesting and implantation.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use.
(i)The following performance characteristics of the cornea following storage in the device must be demonstrated:
(A)Endothelial cell density;
(B)Endothelial cell morphology;
(C)Corneal transparency; and
(D)Central corneal thickness.
(ii)Antimicrobial activity of the device must be demonstrated at the
Free access — add to your briefcase to read the full text and ask questions with AI
21 C.F.R. § 886.4320 (Corneal storage medium with preservatives including antifungals.) — published by Counsel Stack Legal Research, free access to 12M+ legal documents.
Nearby Sections
11
§ 886.4250
Ophthalmic electrolysis unit.§ 886.4270
Intraocular gas.§ 886.4275
Intraocular fluid.§ 886.4280
Intraocular pressure measuring device.§ 886.4300
Intraocular lens guide.§ 886.4335
Operating headlamp.§ 886.4350
Manual ophthalmic surgical instrument.§ 886.4355
Corneal inlay inserter handle.§ 886.4360
Ocular surgery irrigation device.§ 886.4370
Keratome.