§ 880.5405 Active noise attenuation system for infant incubators.
(a)Identification. An active noise attenuation system for infant incubators is a device system that captures environmental noise and outputs noise cancelling acoustic sound waves to attenuate noise in infant incubators in the healthcare environment.
(b)Classification. Class II (special controls). The special controls for this device are:
(1)Non-clinical performance testing under anticipated conditions of use must demonstrate that the device performs as intended, including:
(i)Verification and validation of critical acoustic parameters, including the maximum output of the device;
(ii)Verification and validation of the attenuation performance of the device, including:
(A)Testing with compatible incubator model(s) an