Sadler v. Advanced Bionics, Inc.

929 F. Supp. 2d 670, 2013 WL 898152, 2013 U.S. Dist. LEXIS 32228
CourtDistrict Court, W.D. Kentucky
DecidedMarch 8, 2013
DocketCivil Action No. 3:11-CV-00450-H
StatusPublished
Cited by9 cases

This text of 929 F. Supp. 2d 670 (Sadler v. Advanced Bionics, Inc.) is published on Counsel Stack Legal Research, covering District Court, W.D. Kentucky primary law. Counsel Stack provides free access to over 12 million legal documents including statutes, case law, regulations, and constitutions.

Bluebook
Sadler v. Advanced Bionics, Inc., 929 F. Supp. 2d 670, 2013 WL 898152, 2013 U.S. Dist. LEXIS 32228 (W.D. Ky. 2013).

Opinion

MEMORANDUM OPINION AND ORDER

JOHN G. HEYBURN, II, District Judge.

Breanna Sadler and her parents Michelle and Brian Sadler (collectively, “Plaintiffs”) bring this action to recover for injuries Breanna incurred as a result of the malfunctioning of her cochlear implant against Advanced Bionics, the manufacturer of the device. Advanced Bionics now moves for summary judgment of most of Plaintiffs’ claims on preemption grounds. Many district and circuit courts have addressed medical device preemption issues from a variety of perspectives. As yet, the Sixth Circuit has not contributed any comprehensive discussion on the precise context presented here. The Court has reviewed relevant approaches and decided upon a reasonable one. For the reasons that follow, the Court sustains the motion in part and denies the motion in part.

I.

The facts particular to this case are straightforward. Breanna Sadler suffered from permanent hearing loss until she underwent open-head surgery to receive a cochlear implant on January 16, 2006. Breanna’s cochlear implant device was Advanced Bionics’ HiRes 90k. With the assistance of a cochlear implant, the brain learns to decode noise, and over time, implant recipients are able to distinguish sounds such that they are able to hear again. Although the device is technologically complicated, two principal systems comprise the cochlear implant: an internal component surgically implanted in the skull and an external component that sits outside the ear. The feedthru, which must [674]*674remain hermetic (waterproof) throughout the life of the device, connects these two parts.

In December 2009, a leak in the feedthru that allowed moisture into the device caused Breanna’s cochlear implant to fail, and as a result, she endured an aggressive shock. As Breanna convulsed on the floor, her mother held her down. After Breanna disconnected the device, her mother reconnected it at the urging of a doctor on the telephone. The cochlear implant again shocked Breanna, and Breanna’s mother allowed Breanna to once more disconnect the device. The following day, Brian Sadler took Breanna to an audiologist. While Brian held Breanna down, the doctor reconnected the external processor to Breanna’s skull. For the third time, the device shocked Breanna. When Breanna underwent replacement surgery in February 2010, the Sadlers chose a competitor cochlear implant for their daughter. The lingering physical and emotional effects of the three shocks are disputed.

Breanna’s cochlear implant is subject to significant federal regulation under the Federal Food, Drug, and Cosmetic Act (“FDCA”), which governs medical device manufacturing. The FDCA specifically empowers the Food and Drug Administration (“FDA”) to regulate production and labeling practices of medical product manufacturers and distributors. In amending the FDCA several times since its enactment, Congress expanded the federal role in medical product regulation while preserving the applicability of state laws to the manufacture and distribution of medical products. Wyeth v. Levine, 555 U.S. 555, 566, 129 S.Ct. 1187, 173 L.Ed.2d 51 (2009).

In 1976, Congress enacted the Medical Device Amendment (“MDA”) to broaden the FDCA’s regulatory scope to include medical device manufacturing. Anne-Marie Dega, The Battle Over Medical Device Regulation: Do the Federal Medical Device Amendments Preempt State Tort Law Claims, 27 Loy. U. Chi. L.J. 615, 625-26 (1996). Because medical devices vary in composition and risk, Congress created three classes of these devices with corresponding varied levels of federal administrative control over the covered devices. For Class III devices, such as Advanced Bionics’ cochlear implant, the FDA will grant manufacturers approval to market and distribute their products after obtaining premarket approval (“PMA”).1 PMA “requires that manufacturers conduct extensive testing to obtain clinical results that demonstrate the safety and effectiveness of a device. Both FDA staff members and outside experts must approve the PMA before the FDA will approve the device for commercial use.” Dega, supra, at 627. This involves over 1,000 hours of FDA review. Lohr, 518 U.S. at 478, 116 S.Ct. 2240. Once a device receives PMA approval, the manufacturer must

submit a PMA Supplement for review and approval by FDA before making a change affecting the safety or effectiveness of the device---- [CJhanges for
[675]*675which an applicant shall submit a PMA supplement include ... (6) Changes in the performance or design specifications, circuits, components, ingredients, principle of operation, or physical layout of the device.

21 C.F.R. § 814.39(a) (emphasis added). In sum, federal regulations require Advanced Bionics to obtain PMA approval for the initial cochlear implant device and supplemental approval for any subsequent generation of the device that incorporates changes affecting device safety or effectiveness.

In 1996, Advanced Bionics obtained a PMA for its original cochlear implant, the Clarion Multi-Strategy Cochlear Implant System. Subsequently, Advanced Bionics obtained PMA Supplement # 30 for a later generation of the device, called the HiRes 90k. Advanced Bionics manufactured two versions of the HiRes 90k. First, the PMA Supplement #30 approved the device using a Pacific Aerospace and Electronics, Inc. (“PA & E”) manufactured feedthru (“Vendor A HiRes 90k”). In July of 2003, Advanced Bionics began commercially selling the Vendor A HiRes 90k.

Shortly thereafter, Advanced Bionics began manufacturing the second version of the HiRes 90k; this cochlear implant contained a feedthru manufactured by AstroSeal, Inc. (“Vendor B HiRes 90k”). Advanced Bionics did not seek a PMA Supplement for the Vendor B HiRes 90k. When Breanna underwent cochlear implant surgery in January of 2006, she received a Vendor B HiRes 90k, manufactured in 2005.

A number of issues relating to the HiRes 90k arose prior to Breanna’s implantation surgery. In 2004, Advanced Bionics thoroughly tested failed, explanted cochlear implant devices, and concluded that the devices failed due to excessive moisture concentration in the feedthrus. In September of 2004, Advanced Bionics issued a recall of its unimplanted cochlear implant devices including the HiRes 90k, claiming that moisture sealed into the device during its manufacture caused issues with the proper functioning of the device. After implementing some changes to the device manufacturing process, Advanced Bionics resumed manufacturing and distributing the HiRes 90k. Nevertheless, Advanced Bionics received a warning letter the next year from the FDA informing it of insufficiencies in its operating structure. Because some devices continued to fail, Advanced Bionics issued another recall in March of 2006, which included a recall of unimplanted Vendor B HiRes 90ks. This time, Advanced Bionics concluded that a leak in the feedthru, rather than moisture being sealed in the device during manufacture, caused the excessive moisture levels.

After the 2006 recall, the FDA conducted an on-site investigation of Advanced Bionics’ operations related to the recall.

Free access — add to your briefcase to read the full text and ask questions with AI

Related

Barrios v. Elmore
W.D. Kentucky, 2020
Aquino v. C. R. Bard, Inc.
N.D. Illinois, 2019
Jackson v. E-Z-Go Div. of Textron, Inc.
326 F. Supp. 3d 375 (W.D. Kentucky, 2018)
Moore v. Zydus Pharmaceuticals (USA), Inc.
277 F. Supp. 3d 873 (E.D. Kentucky, 2017)
Spier v. Coloplast Corp.
121 F. Supp. 3d 809 (E.D. Tennessee, 2015)
Hafer v. Medtronic, Inc.
99 F. Supp. 3d 844 (W.D. Tennessee, 2015)
Cline v. Advanced Neuromodulation Systems, Inc.
17 F. Supp. 3d 1275 (N.D. Georgia, 2014)
Eggerling v. Advanced Bionics, LLC
958 F. Supp. 2d 1029 (N.D. Iowa, 2013)

Cite This Page — Counsel Stack

Bluebook (online)
929 F. Supp. 2d 670, 2013 WL 898152, 2013 U.S. Dist. LEXIS 32228, Counsel Stack Legal Research, https://law.counselstack.com/opinion/sadler-v-advanced-bionics-inc-kywd-2013.