Kimberly-Clark Worldwide, Inc. v. Tyco Healthcare Group LP

635 F. Supp. 2d 870, 2009 U.S. Dist. LEXIS 62879, 2009 WL 1940376
CourtDistrict Court, E.D. Wisconsin
DecidedJuly 6, 2009
Docket09-C-429
StatusPublished
Cited by2 cases

This text of 635 F. Supp. 2d 870 (Kimberly-Clark Worldwide, Inc. v. Tyco Healthcare Group LP) is published on Counsel Stack Legal Research, covering District Court, E.D. Wisconsin primary law. Counsel Stack provides free access to over 12 million legal documents including statutes, case law, regulations, and constitutions.

Bluebook
Kimberly-Clark Worldwide, Inc. v. Tyco Healthcare Group LP, 635 F. Supp. 2d 870, 2009 U.S. Dist. LEXIS 62879, 2009 WL 1940376 (E.D. Wis. 2009).

Opinion

DECISION AND ORDER

WILLIAM C. GRIESBACH, District Judge.

This case is presently before me upon Plaintiff Kimberly-Clark’s (“K-C”) motion for a preliminary injunction to prohibit Defendant Tyco Healthcare Group LP (d/ b/a “Covidien”) from marketing and selling a line of endotracheal tubes under the “SealGuard” trademark. K-C alleges that the SealGuard products infringe U.S. Patent No. 6,526,977, the rights to which K-C acquired in 2005 from a German company. The parties have had the opportunity to brief the motion and, on June 25, they appeared for a hearing. For the reasons given below, I conclude that the motion for preliminary relief should be denied.

I. Background

A. The'977 Patent

U.S. Patent No. 6,526,977 describes a product known as an endotracheal cuff, which is a sort of thin balloon that anchors endotracheal and tracheostomy tubes in a patient’s trachea during long periods of intubation (e.g., intensive care). Among the principal purposes of the cuff is to seal out leakage of secretions into the patient’s lungs, which can lead to a fairly common and problematic onset of pneumonia known as ventilator-associated pneumonia (“VAP”). According to the parties, VAP can be deadly but at a minimum it results in longer hospital stays and markedly increased healthcare costs. Generally speaking, endotracheal cuffs prevent this leakage by inflating and expanding to the width of a patient’s trachea, thus forming a *873 tight seal between the breathing tube and the trachea wall.

Some cuffs, known as high pressure, low volume (“HPLV”) cuffs, use the pressure of the inflated cuff to press against the tracheal wall and form a tight seal. In order to achieve an optimum seal, however, the pressure is often too high and can cause damage to the trachea. According to the '977 patent, “[w]hen the filling pressure of the cuffed balloon exceeds the blood flow pressure of the vascular bed supplying the film of the mucous membrane, serious structural lesions of the epithelium might ensue.” (Col. 1:31-34.)

Thus, the purpose of the '977 invention was to create a cuff providing an adequate seal but without the risk of damage arising from high cuff pressure. The invention does so by teaching a cuff of a fixed size that is tailored to the patient’s trachea. This is known as a high volume, low pressure (“HVLP”) cuff. K-C describes the difference as follows: HPLV cuffs are like standard latex balloons, whose size depends on the air pressure inside, whereas the low pressure cuffs are more like Mylar balloons. These will expand until they reach their predetermined and preformed shape, but not any further. The HVLP cuffs are designed to be wider than the patient’s trachea, so that when the cuff is inserted and inflated it will expand until it presses lightly against the trachea and then will fold in upon itself, a process K-C describes as “infolding”.

Although these kinds of cuffs eliminated the problem resulting from high pressure, the creation of folds proved to be a key problem with HVLP cuffs. If the cuff created vertical “draped” folds, these could form channels wide enough to conduct secretions into the patient’s lungs — the very problem cuffs are designed to solve. In other words, the HVLP cuffs are more porous than HPLV cuffs because the folds resulting from infolding create openings for fluids to pass. The '977 patent is intended to alleviate that problem through the design of the folds themselves. The use of an ultra-thin polyurethane material allows the folds that form to be “self-sealing” folds. In other words, the material is so thin that the walls of the folds will press together and either eliminate any channels or openings that might otherwise conduct fluids or cause any “loops” or other openings to be so small that they do not conduct secretions.

By way of crude analogy, imagine an empty cardboard roll designed to hold paper towels or toilet paper. Suppose that you tried to press such a roll into a lead pipe whose circumference was a half-inch narrower than the roll. It is not difficult to imagine the roll bending in on itself and creating a crease or fold. The roll will generally hold its cylindrical shape (apart from the crease) because that is how it is preformed. The fold created, however, will likely be jagged and imperfect, possibly with loops or other openings, an inevitable result of the relatively thick cardboard from which the roll is made. It is not hard to imagine making a similarly shaped roll out of a much thinner material, such as aluminum foil. The thinner the material is, the more complete or perfect the fold will be, because there is little rigidity or excess material to get in the way of the walls pressing together with they fold in upon themselves. And if one used an ultra-thin material and added air pressure to the inside, it is not hard to envision how the walls of the outer and inner pressure would combine to cause the fold walls to press together and essentially seal themselves or remain extremely narrow. The thinner the material, the more likely it is that the folds become tight creases with very narrow loops or other openings.

*874 The '977 patent describes the process as follows:

Surprisingly enough, the flow of secretion can be influenced by a specific design of the cuff folding in the area of the loop of the fold, i.e. at the base of the fold. While in the prior art it has so far been assumed that cuffed balloons with a draped fold cannot rest on the trachea in a sufficiently tight manner because of the low filling pressure, the invention shows a method of inhibiting the flow of secretion, the method being employed in the area of the loop of the fold. When the diameter of the loop is sufficiently small at the base of the fold, the free flow of secretion through the loop is inhibited. When the cuff is blocked, the resultant loops at the deep end of the fold can be reduced with respect to their diameter, for instance by selecting the material or the foil thickness, in such a manner that the flow of secretion is decelerated or, ideally, stopped altogether.

(Col. 2:9-23.)

Claim 1 of the patent sets forth the invention as follows:

A ventilating device for obturating a patient’s trachea as hermetically as possible, comprising a cuffed balloon which 40 blocks the trachea below a patient’s glottis, an air tube, the cuffed balloon being attached to the air tube and being sized to be larger than a tracheal diameter when in a fully inflated state and being made of a sufficiently soft, flexible foil material that forms at least one draped fold in the cuffed 45 balloon when fully inflated in the patient’s trachea, wherein the at least one draped fold formed has a capillary size which arrests free flow of secretions across the balloon by virtue of capillary forces formed within the fold to prevent aspiration of the secretions and subsequent infections related to secretion aspiration.

(Col. 11:39-51.)

The '977 patent issued to Dr. Fred Goebel on March 4, 2003. Dr. Goebel sold the ultra-thin-walled cuffs that embodied his patent through a German company he formed named MicroCuff GmbH. K-C acquired the rights to the '977 patent when it purchased MicroCuff GmbH from Dr. Fred Goebel in 2005. (Decl. of Christopher T. La Testa, Exhs.